Single-use Disposable PPI Samplers are impaction-based samplers that provide a precise match to ISO 7708/CEN size-selection criteria that is specified in the 2016 OSHA final rule on respirable crystalline silica. All Disposable PPI Samplers contain pre-oiled impaction substrate that effectively trap larger particles for sample integrity. Disposable PPI Sampler models are offered for use at 2, 4, or 8 L/min and are available empty or preloaded.
Features
- Collection efficiency precisely matches ISO 7708/CEN criteria adopted in the OSHA final silica rule
- Single use means no cleaning or two-way shipping costs!
- Anti-static plastic
- Selection of flow rates available to meet specific applications
- 8 L/min respirable PPI: Enhances sensitivity (for short-term and/or low concentration sampling) using high flow pumps; ideal for new lower OSHA PEL for silica
- 4 L/min respirable PPI: Enhances sensitivity and can be used with personal pumps; TWA sampling for > 4 hours
- 2 L/min respirable PPI: 8-hour TWA sampling
- Disposable PPI Sampler Options
- Preloaded
- with filter preweighed by SKC
- with filter not preweighed
- Empty for filter loading by a laboratory or the user
- All disposable PPI Samplers are preloaded with pre-oiled impaction substrates in the inlet section
Reusable aluminum PPIs
are available
Performance published in Journal of Physics and referenced in the OSHA final silica rule
Specifications
Performance Profile
| Sampling Rate |
2, 4, or 8 L/min |
| Sample Pump |
AirChek Series for 2 and 4 L/min,
Leland Legacy for
8 L/min |
| Sample Time |
Dependent on method used. Note: SKC tests indicate that a particulate mass of up to 6.8 mg on the four impaction substrates would not affect PPI performance. This amount is equivalent to sampling for 6 hours at 4 L/min in environments where respirable mass concentration is 4.76 mg/m3 and equals 50% of total dust. However, labs have reported to SKC that they prefer no more than 2 mg on the filter for analytical reasons. Therefore, SKC recommends that you work with your lab to determine optimum sample times for your unique sampling conditions. |
| Sample Media |
37-mm, 5.0-µm PVC filter
Models with preweighed filter: filter preweighed at 70 F ± 3 F and 50% ± 5% relative humidity
Use cellulose pad or stainless steel screen for support.
|
| Tubing |
1/4-in ID |
| Impaction Substrate |
Four 3/8-in diameter pre-oiled porous plastic discs |
| Analysis |
Gravimetric or chemical |
| Body Material |
Conductive ABS plastic |
| Dimensions |
Height: (clip to exhaust): 4.25 in
(10.8 cm)
Diameter: 1.8 in (4.6 cm)
Depth: 1.2 in (3.0 cm) |
| Weight |
1.1 oz (31.2 gm) |
| Shelf-life |
18 months from date of manufacture |
Order
| Description |
Cat. No. |
Parallel Particle Impactor (PPI), Disposable, Respirable, Plastic, 8 L/min
Parallel Particle Impactor (PPI), plastic, single-use disposable, four stages, for respirable sampling at 8 L/min. The PPI provides the closest match to the entire ISO 7708/CEN respirable curve compared to any other sampler available, ea. Pre-assembled with four porous plastic disc impaction substrates Requires a 37-mm filter and support purchased separately |
225-384 |
Parallel Particle Impactor (PPI), Disposable, Preloaded, Respirable, Plastic, 8 L/min
Parallel Particle Impactor (PPI), plastic, disposable, four stages, for respirable sampling at 8 L/min. The PPI provides the closest match to the entire ISO 7708/CEN respirable curve compared to any other sampler available. Preloaded with four porous plastic disc impaction substrates, one 37-mm, 5.0-µm PVC collection filter, and one cellulose support, ea |
225-3841 |
Parallel Particle Impactor (PPI), Disposable, Respirable, Plastic, 2 L/min
Parallel Particle Impactor (PPI), plastic, single-use disposable, four-stage, for respirable sampling at 2 L/min. The PPI provides the closest match to the entire ISO 7708/CEN respirable curve, compared to any other sampler available. Pre-assembled with four porous plastic disc impaction substrates, ea Requires a 37-mm filter and support purchased separately, ea |
225-385 |
Parallel Particle Impactor (PPI), Disposable, Preloaded, Respirable, Plastic, 2 L/min
Parallel Particle Impactor (PPI), plastic, single-use disposable, four stages, for respirable sampling at 2 L/min. The PPI provides the closest match to the entire ISO 7708/CEN respirable curve compared to any other sampler available. Preloaded with four porous plastic disc impaction substrates, one 37-mm, 5.0-µm PVC collection filter, and one cellulose support, ea |
225-3851 |
Parallel Particle Impactor (PPI), Disposable, Respirable, Plastic, 4 L/min
Parallel Particle Impactor (PPI), plastic, single-use disposable, four stages, for respirable sampling at 4 L/min. The PPI provides the closest match to the entire ISO 7708/CEN respirable curve, compared to any other sampler available. Pre-assembled with four porous plastic disc impaction substrates, ea Requires a 37-mm filter and support purchased separately |
225-387 |
Parallel Particle Impactor (PPI), Disposable, Preloaded, Respirable, Plastic, 4 L/min
Parallel Particle Impactor (PPI), plastic, single-use disposable, four stages, for respirable sampling at 4 L/min. The PPI provides the closest match to the entire ISO 7708/CEN respirable curve compared to any other sampler available. Preloaded with four porous plastic disc impaction substrates, one 37-mm, 5.0-µm PVC collection filter, and one cellulose support, ea |
225-3871 |
Calibration Adapter for Disposable PPI
SKC Calibration Adapter for the Disposable Plastic Parallel Particle Impactor (PPI) |
225-389 |
Accessories
| Description |
Cat. No. |
Stainless Steel Support Screen, 37 mm
Support Screen, 37 mm, stainless steel, wide mesh |
225-26 |
Cellulose Support Pad, pk/100
Support Pad, cellulose, 37 mm, pk/100 |
225-27 |
Calibration Adapter for Disposable PPI
SKC Calibration Adapter for the Disposable Plastic Parallel Particle Impactor (PPI) |
225-389 |
PVC Filters, 5.0 µm, 37 mm
PVC Filter, 5.0 µm, 37 mm, no support pad included, GLA-5000, pk/100 |
225-5-37 |
Filter-Keeper, 37 mm, with Labels
Filter-Keeper, 37-mm filter storage and archiving system with labels, pk/10 |
225-8303A |
Non-serrated Forceps, Stainless Steel
Forceps, stainless steel, non-serrated flat tips to prevent damage to filter membranes |
225-8371 |