The UMEx 200 Passive Sampler collects sulfur dioxide (SO2) at 15.2 ml/min and nitrogen dioxide (NO2) at 17.3 ml/min. The UMEx 200 contains a tape treated with triethanolamine (TEA) and “blank" to sample from 15 minutes to 24 hours. Analysis is performed by ion chromatography (IC) with conductivity detection.
Features
- Validation range extended below new TLV-TWA for NO2
- Same chemistry as active OSHA Method ID-182
- Accurate, reliable occupational sampling
- Ready to use
- Lightweight sampler clips easily in the breathing zone
- Economical
- No pump or training required
- Low-cost single-use sampler
- Sample medium and blank in one unit
- Safe — no glass or chemical liquids in the sampler
- Validated sampling occupational exposure to sulfur dioxide and/or nitrogen dioxide from 15 minutes to 24 hours
- 3-week sample storage at ambient temperature
Specifications
Performance Profile
Sampling Rate |
SO2: 15.2 ml/min with a
relative standard deviation of 16.5%
NO2: 17.3 ml/min with a relative standard
deviation of 11.4% |
Validated Concentration Range |
SO2: 0.4 to 8 ppm
NO2: 0.051 to 8.5 ppm |
Analysis Method |
Solvent extraction and ion chromatography with conductivity
detection |
Lower Detection Limits |
SO2: 15 minutes: 1.8 ppm
8 hours: 52 ppb
24 hours: 17.4 ppb
NO2: 15 minutes: 0.2 ppm
8 hours: 6.3 ppb
24 hours: 2 ppb
|
Shelf-life |
18 months from date of manufacture at ambient temperature |
Storage |
Before use: Ambient temperature
After use: Analyze within 3 weeks. Can be stored at ambient
temperature or at < 39.2 F (4 C) |
Accuracy
|
SO2: ± 30%
NO2: ± 27% |
Temperature Effects |
SO2: No effect on sampling rate from
20 to 25 C
NO2: No effect on sampling rate from
22 to 40 C
|
Humidity Effects |
No effect from 20 to 80% relative humidity |
Wind Effects |
No effect from 5 to 100 cm/sec |
Interferences |
None identified |
Dimensions |
3.4 x 1.1 x 0.35 in (8.6 x 2.8 x 0.89 cm) |
Weight |
0.38 oz (10.9 gm) |
Slide Cover |
Yellow |
Order
Description |
Cat. No. |
UMEx 200 Passive Samplers, for Sulfur Dioxide and Nitrogen Dioxide, pk/10
UMEx 200 Passive Sampler, for sulfur dioxide and nitrogen dioxide, with validated concentration range of 0.4 to 8 ppm for SO2 and 0.05 to 8 ppm for NO2; single use only, pk/10
Limited shelf-life |
500-200 |
Treated Tape for QC, UMEx 200
Treated Tape for QC, UMEx 200 Treated Tape for QC, UMEx 200, pk/25 |
P20098 |